INSTRUCTIONSSZX7STEREOMICROSCOPE SYSTEMA X 6 6 3 1 Modules described in this manualSZ2-STWHSZ Series EyepiecesObjectivesSZX-EPASZX-ASSZX-STAD1SZH-STA
72Adjusting the Diopter (Zoom Confocality Adjustment)(Fig. 3)#Confirm that the eyepiece clamping knobs @ are tightened firmlybefore proceeding to the
SZX783Using the Eye Shades(Fig. 4)#The eye shades are not provided with ESD protection performance.When ESD protection is required, do not use the eye
95Selecting the Light Path (SZX-TR30)(Fig. 7)Slide the light path selector knob @ to select the desired light path.Light Path Selector Knob Indication
SZX7104-3 Microscope Body1 Zoom Magnification Indication(Figs. 9 & 10)The body zoom magnification is indicated on zoom adjustment knob(right) @.Th
11Microscope body3Using Auxiliary Sleeve SZ2-ET(Fig. 11)As the 0.5X objective has a long working distance, it cannot be used incombination with the st
SZX7124-4 TV Observation and Photomicrography}When TV observation or photomicrography is required, use the SZX-TR30 trinocular tube.A TV camera and/or
13TROUBLESHOOTING GUIDEUnder certain conditions, performance of the microscope may be adversely affected by factors other than defects. Ifproblems occ
SZX714SPECIFICATIONSItem Specifications1. Zoom microscope body. · SZX-ZB7Zoom magnification variation: Left/right optical axis parallel method.Zoom dr
15EyepiecesActualField (mm)OPTICAL PERFORMANCEObjectiveWHSZ10X/10X-H WHSZ15X-H WHSZ20X/20X-H WHSZ30X-HTotalMagnificationDFPL 0.5X-4 4X - 28X 55 - 7.8
SZX716Tools RequiredASSEMBLY8-1 Assembly DiagramThe diagram below shows the sequence of assembly of the various modules. The numbers indicate the orde
178-2 Detailed Assembly Procedures1Installing the Microscope Body(Fig. 13)1. Loosen the body clamping knob @ and insert the microscope body ²gently in
SZX718MountingDismounting5Mounting the Stage Plate(Figs. 16 & 17)1. While applying the stage plate @ to the stage plate holder spring ², fitthe st
199OPTIONAL MODULES9-1 Eye Point Adjuster SZX-EPA1External View and NomenclatureObservation tube mountObservation tube clamping screw2Installation(Fig
SZX7203Using the Aperture Iris Diaphragm}Adjusting the aperture iris diaphragm makes it possible to improve the contrast of the observation image and
213InstallationRotary stageU-SRPMechanical stageU-FMPClamping screwRotary stageU-SRGAllen wrenchClamping screwBX stage adapter type 1SZX-STAD1Mount ho
SZX7229-4 Stage Adapter Type 1 SZH-STAD1The SZH-STAD1 is an adapter providing the similar function to the BX stage adapter type 1 (SZX-STAD1). The dif
23Push up.Clamp.To increaseTo decrease2External View and NomenclatureClamping screw holesStageLoad capacity: 1 kgClamping knob Focus adjustment kn
SZX724Bring to the centerof the stroke. Adjusting the focus (Fig. 21)1. Rotate the focus adjustment knob @ of the SZ2-FO up/down movingstage to bring
252InstallationStage plate mount holeFriction surfacesStage plateUse the one providedwith the stand.Gliding stageStage plate mount holeApplicable stan
SZX726#Before mounting, remove dirt and dust from the mount surfaces and handle carefully so as not to damage them.Stage plateFriction surfacesStage p
CONTENTSIMPORTANT — Be sure to read this chapter for safe use of the equipment. —1 NOMENCLATURE31-2Correct assembly and adjustments are critical f
27Place a specimen on the stage plate, hold the cup stage by the edge,and tilt the cup stage slowly. (Fig. 22)}If the specimen slips on the stage plat
Shinjuku Monolith, 3-1, Nishi Shinjuku 2-chome, Shinjuku-ku, Tokyo, JapanWendenstraße 14-18, 20097 Hamburg, Germany3500 Corporate Parkway, P.O. Box 61
1IMPORTANTSAFETY PRECAUTIONSThe SZX7 stereomicroscope is provided with the ESD (Electro-Static Discharge) protection specifications. It is providedwit
SZX723CautionIf the microscope is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition,the equipment ma
3NOMENCLATURE * Modules marked * are given the ESD specifications.(Note) The modules shown in the following diagram are merely the typical examples. F
SZX74CONTROLSEye shades (Page 8)Eyepiece clamping knobsBody clamping knobThreaded for accessorymountingAccessory mounting screws(x 5)Stage plateAllen
5OBSERVATION PROCEDURE}If you have not assembled the microscope yet, please read Chapter 8, “ASSEMBLY” (pages 16 to 18).3-1 Preparation(Page)1. Confir
SZX76OPERATION4-1 Stand1Using the Stage1. When reflected light illumination is used, the stage plate can be placedeither the white or black surface fa