Olympus AU2700 User Manual Page 22

  • Download
  • Add to my manuals
  • Print
  • Page
    / 65
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 21
- 22 -
(3.6) Others
(1) Contents and formats of text.
Item Digit Contents Remarks
1. Rack No.
(Optional)
4 or 5 '0001'-'9999' or
'00001'-'99999'
STAT sample is space
When Rack no. is no setting,
Rack No. is '00000'.
The digit of rack no.
and cup pos. and
these using is
changeable in
[PARAMETER]
[ONLINE]
[PROTOCOL]
menu.
In case ,receive
T.R.I.
in batch.
2. Cup position
(Optional)
2 rack : '01'-'10'
STAT : '01'-'22'
This is automatically
assigned when rack
No. is in use.
3. Sample type 1 space : serum
U : urine
X : other
4. Sample No. 4 '0001'-'9999' : Routine sample
'E001'-'E999' : Emergency sample
'P001'-'P999' : Stat sample
'R001'-'R999' : Reagent blank
'A001'-'A999' : Calibration sample
'Q001'-'Q999' : QC sample
5. Sample ID 4-26 Number or character
Number of digits can
be changed in the
[PARAMETER]
[FORMAT]
[FORMAT
REQUISITION]
menu.
6. Original
Sample No.
4 '0001'-'9999' : normal sample
'E001'-'E999' : emergency sample
'P001'-'P999' : stat sample
For Repeat Run Results.
7. Reserved 4 Space
8. Block 1 '0'-'9' Last text is 'E'
Identification
No.