Olympus AU2700 User Manual Page 40

  • Download
  • Add to my manuals
  • Print
  • Page
    / 65
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 39
- 40 -
(5.2) Sending test result protocol
(1) Transfer sequence for one session.
General sequence :
A. Example 1
Shift next Dx-transfer when Online Error occurred.
AU2700/5400
DB Dx
0
Dx
E
Dx
0
Dx
E
Dx
E
DE
2 blocks
Host
B. Example 2
Stop session when Online Error occurred.
AU2700/5400
DB Dx
0
Dx
E
Dx
0
Dx
E
Dx
E
DE
Host
(2) Detail
Data
Trans. Kind of text Timing/condition Normal end Online error
Real DB (START) This text is transmitted when It is shifting the It is output the
-time it is shifted from stand-by transfer of the alarm and it is
mode to measure mode and just test results. executed the below
behind the first Dx is transmitted.
Error protocol is
Dx (result data) It is transmitted one by one It is shifting next stop.
the samples that are finished The transfer of It is canceled the
measuring. Dx/DE. transfer of Dx/DE
Error protocol is
continue
It is shifting next
the transfer of
Dx/DE.
DE (end) This text is transmitted after It is ended the It is output the
the last transfer of Dx text session of transfer of alarm and ended
and the below shifting mode. The text of test the session of the
results. transfer of the
1) measure mode to standby text of test results.
2) measure mode to stop
This text is transmitted when
the communication is canceled
by the online alarm.
Page view 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 64 65

Comments to this Manuals

No comments