Olympus AU2700 User Manual Page 57

  • Download
  • Add to my manuals
  • Print
  • Page
    / 65
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 56
- 57 -
4)ONLINE ILLEGAL TEXT CODE [ aa ]
<Operation>
(1).Disregard the test requisition information text for the sample generating the alarm.
(2).Continue to receive the test requisition information text for samples following the alarm.
<Description>
(1).When receiving the test requisition information text using online, the text classification code is out of specification.
(2).The display of [ ] and its meaning are as follows.
aa :
----------------------------------------------------------------
R_ : Normal sample
RH : Repeat run sample
----------------------------------------------------------------
5)ONLINE ILLEGAL TEXT BLOCK No. [ aa -> bb ]
<Operation>
(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP",
the following events occur:
1).Disregard the test requisition information text for the sample generating the alarm.
2).System will not receive text for samples following the sample generating the alarm.
3).In a case when the text requisition is received in real-time, the "STOP" is ignored and the text requisition text
will be received with real-time again during the next analysis.
(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue",
the following events occur:
1).Disregard the test requisition information text for the sample generating the alarm.
2).The test requisition information text for the next sample will be received.
<Description>
(1).When receiving the requisition information text using online, the block identification No. is out of specification.
(2).The display of [ ] and its meaning are as follows.
aa : Previously received block identification No.
bb : Last received block identification No.