Problem
Cause Remedy Page
g) Image is blurred. Objective is engaged incorrectly in the
light path.
Make sure that revolving nosepiece
clicks into place correctly.
Specimen is tilted with respect to the
stage.
Place the specimen correctly on the
stage and secure it with the specimen
holder.
16
h) Field of view of one eye does not
match that of the other.
The interpupillary distance is incorrect.
Adjust the interpupillary distance. 18
Incorrect diopter adjustment. Adjust the diopter. 18
You are not accustomed to parallel opti-
cal axis.
When looking into eyepieces, do not
stare at image from the beginning but
see the overall field of view. It is some-
times recommended to turn your eyes
away from the eyepieces, look far off
and look into the eyepieces again.